Fume Hood Selection, Use and Face-Velocity Checks

November 24, 2026

Overview

A fume hood is a local exhaust device that captures airborne chemical contaminants at the point where they are generated. Its performance depends on hood type, room air supply, placement and the way the sash and materials are handled. Face-velocity checks provide evidence that the hood continues to operate as designed.

Principles

Air is drawn through the opening toward the interior and discharged outside the laboratory or into a treatment system. Capture occurs when inward airflow at the face opposes room cross-drafts and the momentum of the work. Face velocity is the average air speed at the hood opening. It is measured with a calibrated instrument at defined grid points and compared with the value recorded at commissioning. One reading at the centre of the opening is not representative. Smoke visualisation tests show whether short-circuiting or eddies develop near the sash or the work surface.

Selection criteria

Hood type Where it fits Limitation
Conventional constant-volume hood Routine work with fluctuating sash position where room air supply is stable Higher air consumption than variable-volume systems
Variable-air-volume hood Laboratories with controlled supply and a need to reduce air changes Performance depends on a functioning control system
Ducted hood with dedicated construction Perchloric acid or radioactive work specified by the risk assessment Requires a dedicated exhaust and washing or monitoring arrangement
Filtered recirculating hood Specific vapours for which the filter is approved and monitored Not a substitute for ducted exhaust for uncharacterised mixtures

Face-velocity check procedure

  1. Confirm that the measuring instrument has a current calibration and record hood identifier, date and sash height.
  2. Set the sash to the working height specified in the operating procedure; remove materials that obstruct the baffle.
  3. Allow the room ventilation to reach steady conditions and note open doors or nearby equipment that disturb the airflow.
  4. Measure at each defined grid point across the opening and calculate the average; record individual values as well as the average.
  5. Compare the average with the commissioning value and the range specified by the institutional standard; record any deviation.
  6. Perform a smoke visualisation test at the sash and inside the hood, and document the behaviour of the plume.
  7. Label the hood with the result and date. Remove a hood from service when results fall outside the specified range and investigate the cause.

Troubleshooting

Symptom Likely cause Check Corrective action
Face velocity below the specified range Blocked baffle or reduced exhaust flow Inspect baffle, duct and fan status Clear the obstruction and repeat the measurement
Smoke escapes past the sash Cross-draft, high sash or work near the front edge Observe room airflow and work position Lower the sash, move work inward and reassess the room balance
Large variation between grid points Obstructed air entry or poor baffle condition Review stored items and internal layout Remove stored chemicals and restore the baffle
Odour in the laboratory during use Inadequate capture or exhaust discharge near an air intake Trace the plume and inspect discharge location Correct the work position and have the exhaust system evaluated

Safety, ethics and limitations

A fume hood does not protect against every hazard. Highly reactive substances, toxic powders and biological agents may require a different enclosure or additional controls defined in the risk assessment. Chemicals stored in the hood reduce the usable working area and can block the baffle. The hood does not replace personal protective equipment. A face-velocity value alone does not prove containment; visualisation and, where required, a formal containment test are needed to interpret the result.

Key takeaways

  • Hood type is selected from the chemical hazard and the exhaust arrangement.
  • Face velocity is an average of defined grid points, not a single centre reading.
  • Smoke visualisation complements the numerical measurement.
  • A hood outside its specified range is taken out of service pending investigation.

References

  1. Occupational Safety and Health Administration. Occupational Exposure to Hazardous Chemicals in Laboratories, 29 CFR 1910.1450.
  2. American National Standards Institute and American Industrial Hygiene Association. ANSI/AIHA Z9.5, Laboratory Ventilation.
  3. American Society of Heating, Refrigerating and Air-Conditioning Engineers. ANSI/ASHRAE Standard 110, Method of Testing Performance of Laboratory Fume Hoods.